What belongs in a non-GLP report that has to support an IND
A non-GLP study is not an unregulated one. The difference is the paperwork that makes a number traceable — and its absence is the usual reason a study gets repeated.
Non-GLP describes the study, not the standard
The phrase is routinely read as permission to skip documentation. It is not. GLP is a formal quality system covering facilities, personnel, archives and independent quality assurance; a non-GLP study sits outside that system but still has to produce data a reviewer can follow.
For an IND submission the question is not "was this GLP?" but "can someone who did not run the experiment reconstruct what was done, and reproduce the reported number?" A study that cannot answer that is not cheaper — it is a study that gets repeated at the worst possible moment.
There are two practical consequences. The documentation load sits at the end of the programme, not the beginning: the report is where a year of plates either becomes evidence or stays a set of files. And the reporting template matters as much as the instrument list, which is why it belongs in the scoping conversation rather than in the final email.
The five parts a reviewer looks for
A report that survives review tends to have the same five parts, whatever the assay was.
- Study plan — objective, test system, controls, acceptance criteria, replicate structure, concentration range and analysis plan.
- Methods — platform, instrument settings, reagents and lot numbers, buffers, plate maps, software and version.
- Raw data — instrument-native files, exports and integration parameters, with a naming scheme that links each one to a result.
- Analysis — curve model, weighting, outlier rule, number of replicates, and the reported value with its interval.
- Conclusions and limits — what the data supports, and what it explicitly does not.
Most deficient reports are not missing the methods or the conclusions. They are missing the middle: the link between a number in a table and the acquisition that produced it.
What makes each part usable
The study plan is where acceptance criteria are fixed, before the first plate. Signal-to-background, Z′, control performance and coefficient of variation are agreed in writing, and a run that misses them is either repeated or reported as a failed run. Criteria written after the results are known are not criteria.
Methods should be specific enough to re-run the assay without asking a question. For a plate-based readout that means detection chemistry, plate type, dispense volumes, incubation times and reader settings. For acoustic dispensing it means the source plate format, transfer volumes and any backfill or normalization step. For SPR it means chip type, immobilization chemistry, contact time, flow rate, regeneration conditions and reference-channel handling.
Raw data should be delivered, not summarised. The distinction matters when a reviewer asks why one concentration point sits off the curve: without the acquisition file and the integration settings, the only possible answer is to take it on trust.
Analysis needs to declare its rules in advance. A four-parameter fit on unweighted data, an outlier rule applied after inspection, a single replicate reported without an interval — each is defensible only if it was stated before the data arrived.
Limits are part of the result. A report that says a compound is inactive below the assay’s solubility limit is more useful than one claiming inactivity without qualification.
The traceability chain
The single test of a report is whether a reported value can be walked backwards: from the table value to the analysis file, to the integration or fit settings, to the raw acquisition, to the run date and the operator who ran it.
- Table value → analysis file → fit or integration settings → raw acquisition → run date and operator
- Every step should be a file, not a verbal explanation
If any step in that chain is a conversation rather than an artefact, the report is one question away from failing. Building the chain costs almost nothing during the study, and a re-run after it.
Six gaps that trigger a repeat study
- Acceptance criteria never written down, so a marginal run cannot be judged.
- Reagents recorded by name but not by lot, making a shift in performance unattributable.
- Single-replicate potency values reported without an interval.
- Curve fits with no statement of model, weighting or constraints.
- Raw instrument files withheld, so unusual points cannot be examined.
- Reference compound run once, so plate-to-plate drift has no anchor.
None of these are analytical failures. All of them are reporting failures, and a reviewer finds them in a few minutes.
How this runs at AXK Bio
Every service line is reported against the same documentation standard, so a programme that moves from enzyme assays into Biacore SPR does not restart its paperwork. Raw instrument files are delivered alongside the processed tables, and the report states the fit model it used together with the limitations that model implies.
What to ask before commissioning a study
These five questions separate a report from a data dump, and they can be asked before any purchase order exists.
- Will the study plan state acceptance criteria before work begins?
- Are raw instrument files included, or available on request?
- Which curve model and weighting will be used, and is that stated in the report?
- How are failed runs handled and reported?
- Can a reported value be traced back to a specific acquisition?
A laboratory that answers these without hesitation is not necessarily better at the bench. It is, however, more likely to hand over something that does not need to be repeated.